To eliminate the cross-contamination.
Protect your patient and your testing instruments with the Pulmonary Function Filter. provides a high degree of protection for the patient as well as your testing equipment, while its low resistance allows for accurate testing results.Independent laboratory tests have verified that this pulmonary function filter is highly effective. consistently filters out 99.9% of bacteria and 99.9% of viruses.
According to ATS recommendations for flow resistance in pulmonary function instrumentation, which suggest Resistance should be below 1.5 cm H2O/L/sec at Flow rates less than 12 L/sec1. The easily surpasses that recommendation, having a measured resistance of less than 0.78 cm H2O/L/sec at Flow rates of 12 L/sec.
1.Am Rev Respir Dis(1987)136:1285-1298
{ WARNING: The use of this filter and mouthpiece (if included) is limited to a single patient, single session. Not intended for reprocessing. It is not to be used with neonate patients. Do not soak, rinse, wash, gas, or steam sterilize this Product. Patient secretions and wetting with cleaning solutions or liquid detergents may degrade the resistance and effectiveness of this filter.CAUTION: Federal law(USA) restricts this device to sale by or on the order of a physician. }
|